Detail Mesin Anestesi ACOMA VIGOR 21DX
Specification
• Power supply ( only for DX type)
Rated power voltage : AC100-240V ( as specified)
Power frequency : 50/ 60Hz
Power input : 7VA
Type of protection against electric shock : Type : Class I
Level : B type equipment
Auxiliary power outlet : Four; Maximum rated amperage 5A ( w/ Circuit breaker)
• Anesthetic circuit
Gas supply pressure : 350~ 500kPa ( common to O2, N20)
Flowmeter : Flow ranee continuously variable; 0- 10 L/ min.
( common to O2, N2O)
( DXonly... w/ pressure gauge for each gas, 0-10 x 100kPa)
Canister ( CO2 absorption device) capacity : 1, 340mL ( 1, 200g)
Oxygen flush : 35~ 75L/ min.
Auxiliary cylinder yoke : O2 only ( one) ; ( DX onh) w/ pressure gauge ( 0-200 x 100kPa)
• Monitors
Circuit pressure gauge : -10 - 70hPa
Oxygen monitor ( DX only) : Galvanic cell measuring system; Button switch
ON/ OFF, and ON by microswitch ( as soon as O2 gas flow
rises to over 1.8L/ min.' . monitor sets to ON)
Measuring ranse/ aceuracy : 0~ 99% O2; 15-25% ... ± 1 % 02; 25-100% ...
± 3% 02
Calibration : 21% calibration
O2 sensor life : 900.000% H
# Safety devices/ Alarms
Ensured minimum O2 flow : Controlled so that when O2 gas flow is less than ' 1.8 L/ min.' .
N2O gas won' t flow.
O2 gas supply pressure : If pressure drops below ' 240~ 260kPa' , whistle sounds.
Below ' 170-21 OkPa' , whistle' sounds at the same tune as
N2O supply is shut off.
Insufficient O2 cone' alarm ( DX only) : O2 cone, selectable at ' 25% ' or ' 18% ' ,
and if cone, goes below selected value, alarm activates.
( The ' O2 cone. indicator display changes from green to
blinking in red. and alarm sounds.)
Calibration unavailable' alarm ( DX only) : When O2 sensor is dead or defective, green' --' mark is
displayed in the O2 cone, indicator, and calibration
unavailable' ' indicator blinks as alarm sounds.
O2 sensor cable disconnected' alarm : Activates if the cable is disconnected or damaged.
( DX only) ( The red' — ' mark blinks on the O2 cone. indicator display,
and alarm sounds.)
Alarm pause ( DX only) : Usable with die following alarms:
( 1) O2 cone, below ' 25% '
( ' Pause' not available when alarm activates at ' 18% ' level.) ;
( 2) Calibration unavailable; and,
( 3) O2 sensor cable disconnected.
* During the pause, the indicator blinks, and once the condition
has beencorrected, alarm is automatically cancelled.
# Dimensions/ weight
Outer dimensions... : 486( w) x 582( d) x 1, 360( h) mm;
Monitor shelf... : 410( w) x 300 ( d) mm, approx. 20kg monitors, etc. mountable;
Table top... : 480( w) x220( d) mm at 880mm height;
Main unit weight... : DX - approx. 70kgs.,
ST- approx. 65kgs.
# Usage conditions
Room temperature... : 10~ 40° C:
Relative humidity... : 30-85% ;
Atmospheric pressure... : 700-1, 060hPa